What is an HPLC Safety Cap?
An HPLC Safety Cap is a technically coordinated closure for the eluent bottle of an HPLC system. It seals the bottle opening and the capillary feedthroughs, guides the supply capillaries safely through the closure and integrates controlled pressure equalisation via a vent valve. Together with fittings, blind plugs and inlet filters, it forms a controlled closed solvent supply system. This reduces uncontrolled air and vapour exchange and helps preserve the prepared composition of the mobile phase.
Why do I need an HPLC Safety Cap?
An HPLC Safety Cap reduces solvent vapours and evaporation losses. It also protects the eluent from airborne particles and supports orderly, safe and reproducible capillary routing.
Which criteria are used to select an HPLC Safety Cap?
Important selection criteria include: Analytical or preparative application; Isocratic or gradient method; Number of eluent bottles; Bottle material and bottle thread; Number of supply capillaries; Capillary outside and inside diameters; Required flow rate; Eluents and additives used; PFAS sensitivity of the application; Possible future extensions The application determines the Safety Cap configuration – not the other way round.
Which Safety Cap size do I need?
The size depends on the maximum number of supply capillaries required. Typical variants are: Safety Cap I: one port; Safety Cap II: two ports; Safety Cap III: three ports; Safety Cap IV: four ports; Safety Cap VI: six ports Unused ports must be closed with suitable blind plugs.
Why are several fitting sizes supplied with a Safety Cap?
For analytical Safety Caps, SCAT supplies fittings for the common capillary outside diameters of 1.6 mm, 2.3 mm and 3.2 mm for each capillary port. This allows the appropriate fitting to be selected during installation. The Safety Cap does not have to be specified for a single capillary size when ordering.
What are blind plugs used for?
An open port creates a direct connection between the eluent bottle and the laboratory environment. This can allow solvent vapours to escape while air, moisture and contaminants can enter the bottle. At the same time, the composition of volatile eluent mixtures may change. The following therefore applies to standard capillary ports: occupied ports are sealed with the appropriate fitting, while unused ports are closed with a blind plug.
Which Safety Cap fits a GL 45 bottle?
A Safety Cap with a matching GL 45 internal thread is required for a GL 45 bottle. In addition, the number of capillary ports, capillary diameters, vent valve, application type and, where applicable, requirements for PFAS-sensitive analyses must be considered.
Why must a vent valve be replaced after six months?
SCAT specifies a service life of six months for standard and GreenLine vent valves. After this period, the complete valve is replaced preventively, even if no visible changes can be detected. Earlier replacement is required if the valve is damaged, heavily contaminated, wetted or flooded with eluent, or if pressure equalisation no longer functions reliably. The replacement date should be documented directly on the valve or in the maintenance plan.