Air Valve
Air valve (Belüftungsventil): Lets the required air into the eluent bottle while eluent is withdrawn.
What is the function of a vent valve?
During operation, the HPLC continuously withdraws eluent from the bottle. Without pressure equalisation, negative pressure would develop, which can impair delivery and promote the formation of air bubbles. The vent valve opens when required, allows filtered ambient air to enter and closes again after pressure equalisation. A permanently open bottle is not an equivalent substitute because solvent vapours can escape and ambient air, moisture or particles can enter in an uncontrolled manner.
Further information: Vent valves for HPLC eluent bottles
How does a SCAT vent valve work?
The vent valve contains a filter membrane and a non-return valve, also known as a duckbill valve.
The filter membrane retains particles from the ambient air. Under negative pressure, the valve opens towards the inside of the bottle and closes again after pressure equalisation.
What is the difference between the standard and GreenLine vent valves?
The standard vent valve has a PTFE filter membrane and is intended for conventional analytical and preparative HPLC applications.
The GreenLine vent valve uses a fluorine-free filter membrane and is part of a coordinated PFAS-sensitive supply system.
What thread do SCAT vent valves have?
SCAT-Vent valvee besitzen ein Gewinde mit der Bezeichnung UNF 1/4" 28G.
The valve is screwed into the designated port of a compatible Safety Cap.
What is the Luer-Lock connection on the vent valve used for?
The Luer-Lock connection enables special applications, for example:
- Purging the eluent with gas
- Protecting oxidation-sensitive eluents
- Connecting a drying tube
- Reducing moisture ingress
Suitability depends on the application, the medium and the operational requirements.
Why must a vent valve be replaced after six months?
SCAT specifies a service life of six months for standard and GreenLine vent valves. After this period, the complete valve is replaced preventively, even if no visible changes can be detected. Earlier replacement is required if the valve is damaged, heavily contaminated, wetted or flooded with eluent, or if pressure equalisation no longer functions reliably. The replacement date should be documented directly on the valve or in the maintenance plan.