Suction Filter

Suction filter (inlet filter, solvent filter): Particle filter at the end of the supply tubing.

Suction filters in the shop

What is the function of an inlet filter?

The inlet filter is located at the lower end of the supply capillary.

It retains particles, precipitates and solid contaminants before they enter the capillary and the HPLC system.

Further information: Inlet filters for HPLC eluent bottles

What is the difference between an inlet filter and a vent valve?

The vent valve filters the ambient air entering the bottle.

The inlet filter filters the liquid eluent before it enters the supply capillary.

Which inlet filters does SCAT offer for conventional and PFAS-sensitive applications?

For conventional applications, an inlet filter made of PFA/PTFE with a pore size of 5 µm is available.

For PFAS-sensitive systems, a fluorine-free version made of UHMW-PE with a pore size of 20 µm is used. The different pore size must be taken into account in system design and priming.

Is a smaller pore size always better for inlet filters?

No.

A smaller pore size retains finer particles but can increase flow resistance and clog more quickly. Filter area, filter geometry, viscosity and flow rate must be considered together.

When must an inlet filter be replaced?

Replacement is required in the event of:

  • Visible deposits
  • Discolouration
  • Damage
  • Restricted flow rate
  • Recurring air bubbles
  • Unusual suction behaviour
  • Insufficient hold on the capillary

There is no generally applicable purely time-based replacement interval for inlet filters.

Related terms

Air valve, Supply tubing, HDPE. All terms in the HPLC glossary, more answers in the FAQ.