Retention Time

Retention time (Retentionszeit): Time between sample injection and detection of a component at the detector.

Air valves in the shop

Why is eluent quality important for HPLC analyses?

Contamination or evaporation-related changes in the mobile phase can affect chromatographic results.

Possible consequences include changes in retention times, unstable baselines, additional peaks and reduced reproducibility. A Safety Cap does not clean the eluent, but it helps preserve its prepared quality and composition as far as possible.

What effect do isocratic and gradient methods have?

An isocratic method uses a single eluent or a constant solvent mixture.

A gradient method uses two or more eluents from separate bottles. The individual eluent paths should be clearly labelled A, B, C or D.

What are typical causes of air bubbles in the suction line?

Air or gas bubbles can enter the suction capillary or form within the solvent supply system. Possible causes include an inlet filter that is not fully immersed, an insufficient fill level, a kinked or damaged capillary, a clogged inlet filter, a vent valve with restricted function, or incomplete venting after a bottle change. Air can also be drawn into the system through leaks on the suction side without liquid necessarily leaking out. The entire solvent supply path from the bottle to the pump inlet should therefore first be checked systematically. If the problem persists, pump-side and instrument-side causes must also be considered.

Why must a vent valve be replaced after six months?

SCAT specifies a service life of six months for standard and GreenLine vent valves. After this period, the complete valve is replaced preventively, even if no visible changes can be detected. Earlier replacement is required if the valve is damaged, heavily contaminated, wetted or flooded with eluent, or if pressure equalisation no longer functions reliably. The replacement date should be documented directly on the valve or in the maintenance plan.

Related terms

Eluent, Air valve, Suction filter. All terms in the HPLC glossary, more answers in the FAQ.