Fitting and Ferrule
Fitting: Screw and fastening element for a line. Ferrule: Sealing element of a fitting.
What is the difference between the outside and inside diameter of a capillary?
The outside diameter determines which fitting can mechanically secure and seal the capillary.
The inside diameter influences flow resistance, line volume, flushing time and possible flow rate.
Which fitting sizes are available?
For analytical applications, fittings are available for capillary outside diameters of 1.6 mm, 2.3 mm and 3.2 mm.
For preparative applications, depending on the design, ports for 4.0 mm, 4.76 mm and 6.35 mm are also available.
Further information: HPLC capillaries and fittings
What do the fitting colours mean?
The colour coding makes it easier to assign the fittings:
- Green: 1.6 mm
- Purple: 2.3 mm
- Blue: 3.2 mm
However, the colour does not replace checking the actual capillary outside diameter.
What is the difference between a fitting and a ferrule?
The fitting is the screw and fastening element.
The ferrule is the sealing element that is pressed against the capillary and sealing surface when tightened. In many SCAT fittings, the ferrule is already integrated.
Why are several fitting sizes supplied with a Safety Cap?
For analytical Safety Caps, SCAT supplies fittings for the common capillary outside diameters of 1.6 mm, 2.3 mm and 3.2 mm for each capillary port.
This allows the appropriate fitting to be selected during installation. The Safety Cap does not have to be specified for a single capillary size when ordering.
Related terms
One-piece fitting, Capillary tubing, Blind plug. All terms in the HPLC glossary, more answers in the FAQ.